Camera-Derived Photoplethysmography (rPPG) and Speckle Plethysmography (rSPG): Comparing Reflective and Transmissive Mode at Various Integration Times Using LEDs and Lasers

Sensors (Basel). 2022 Aug 13;22(16):6059. doi: 10.3390/s22166059.

Abstract

Background: Although both speckle plethysmography (SPG) and photoplethysmography (PPG) examine pulsatile changes in the vasculature using opto-electronics, PPG has a long history, whereas SPG is relatively new and less explored. The aim of this study was to compare the effects of integration time and light-source coherence on signal quality and waveform morphology for reflective and transmissive rSPG and rPPG. Methods: (A) Using time-domain multiplexing, we illuminated 10 human index fingers with pulsed lasers versus LEDs (both at 639 and 850 nm), in transmissive versus reflective mode. A synchronized camera (Basler acA2000-340 km, 25 cm distance, 200 fps) captured and demultiplexed four video channels (50 fps/channel) in four stages defined by illumination mode. From all video channels, we derived rPPG and rSPG, and applied a signal quality index (SQI, scale: Good > 0.95; Medium 0.95−0.85; Low 0.85−0.8; Negligible < 0.8); (B) For transmission videos only, we additionally calculated the intensity threshold area (ITA), as the area of the imaging exceeding a certain intensity value and used linear regression analysis to understand unexpected similarities between rPPG and rSPG. Results: All mean SQI-values. Reflective mode: Laser-rSPG > 0.965, LED-rSPG < 0.78, rPPG < 0.845. Transmissive mode: 0.853−0.989 for rSPG and rPPG at all illumination settings. Coherent mode: Reflective rSPG > 0.951, reflective rPPG < 0.740, transmissive rSPG and rPPG 0.990−0.898. Incoherent mode: Reflective all <0.798 and transmissive all 0.92−0.987. Linear regressions revealed similar R2 values of rPPG with rSPG (R2 = 0.99) and ITA (R2 = 0.98); Discussion: Laser-rSPG and LED-rPPG produced different waveforms in reflection, but not in transmission. We created the concept of ITA to investigate this behavior. Conclusions: Reflective Laser-SPG truly originated from coherence. Transmissive Laser-rSPG showed a loss of speckles, accompanied by waveform changes towards rPPG. Diffuse spatial intensity modulation polluted spatial-mode SPG.

Keywords: PPG; SPG; camera-based; laser speckle; optical monitoring; speckle contrast analysis.

MeSH terms

  • Humans
  • Lasers*
  • Photoplethysmography* / methods

Grants and funding

This research received no external funding.