Purpose: To evaluate mechanical damage of simulated lenticular debris on corneal endothelial cells during phacoemulsification.
Setting: University setting, Salt Lake City, Utah.
Design: Experimental study.
Methods: 12 New Zealand rabbits underwent bilateral surgery. After incision, 10 1.0 mm beads were injected into the anterior chamber of test eyes to simulate lenticular debris. A balanced salt solution was injected into the contralateral eyes with the same injector. The Intrepid Balanced Tip of the Centurion Ozil handpiece was inserted into the anterior chamber, and 60% torsional ultrasound was applied for 20 seconds with 50 mL/min flow, 70 mm Hg intraocular pressure, and 600 mm Hg vacuum. Cumulative dissipated energy (CDE) was noted. After euthanasia and enucleation, the corneas were removed, stained with trypan blue/alizarin red, and photographed in a standardized manner (×400 photographs from 5 specific areas and 1 overview photograph from each corneal button). The ImageJ program was used to evaluate cell damage and loss in the photographs.
Results: CDE was 4.64 ± 0.19 for test eyes with beads and 4.66 ± 0.17 for control eyes without beads ( P = .58). The percentage of intact cells was statistically higher in the control group ( P = .0005), whereas the percentages of damaged and lost cells were statistically higher in the test group ( P = .0010 and .0005, respectively). This was observed in the analysis of the ×400 photographs and overview photographs.
Conclusions: This study suggests that lenticular debris swirling around the anterior chamber may cause significant damage to the corneal endothelium, providing further insight on mechanisms of damage during phacoemulsification.
Copyright © 2022 Published by Wolters Kluwer on behalf of ASCRS and ESCRS.