Birdcage coils are widely used in preclinical MRI as they perform well, allow for quadrature drive, and can provide a homogeneous transmit field. Unlike in larger bore scanners, an RF shield is essential to avoid strong cross-talk with gradient coils that are in close proximity. However, gradient switching induces eddy currents that heat the shield and coil and impair the temporal signal-to-noise ratio (tSNR). The motivation of this study is to investigate the performance of different designs of RF shields on a birdcage coil used for high resolution functional MRI of small primates at 9.4 T. We found the choice of materials for RF shields significantly affected ghosting and tSNR in fMRI scans. Both ultrathin foils and a slotted pattern reduce eddy currents and improve imaging quality. Our results also demonstrate that a 9-um-thick copper foil is sufficiently thin to reduce the eddy current effects for high-resolution fMRI scans and there is no need for high-cost 4-um-thick foil. For slotted shields, our results demonstrate that the number of slots should be carefully considered, and an excessive number of slots can lead to a lower SNR and tSNR. We believe the results from this study can be used as a reference to design future RF coil shields selection for preclinical scanners.
Keywords: Birdcage coil; EPI; Eddy current; RF shielding; fMRI.
Copyright © 2022 Elsevier Inc. All rights reserved.