Background: No detailed data on left bundle branch block (LBBB) and permanent pacemaker implantation (PPI) exist from randomised clinical trials comparing the ACURATE neo and CoreValve Evolut devices.
Aims: Our aim was to assess the incidence and impact of new LBBB and PPI with self-expanding prostheses from a powered randomised comparison.
Methods: From the SCOPE 2 trial, 648 patients with no previous pacemaker were analysed for PPI at 30 days, and 426 patients without previous LBBB were adopted for analysis of LBBB at 30 days. Results: At 30 days, 16.5% of patients required PPI; rates were higher in CoreValve Evolut compared to ACURATE neo recipients (21.0% vs 12.3%; p=0.004). Previous right bundle branch block (odds ratio [OR] 6.11, 95% confidence interval [CI]: 3.19-11.73; p<0.001) was associated with an increased risk of PPI at 30 days, whereas the use of the ACURATE neo (OR 0.50, 95% CI: 0.31-0.81; p=0.005) was associated with a decreased risk. One-year mortality was similar in patients with and without new PPI. A total of 9.4% of patients developed persistent LBBB at 30 days, with higher incidences in CoreValve Evolut recipients (13.4% vs 5.5%; p=0.007). New LBBB at 30 days was associated with lower ejection fraction at 1 year (65.7%±11.0 vs 69.1%±7.6; p=0.041).
Conclusions: New LBBB and PPI rates were lower in ACURATE neo compared to CoreValve Evolut recipients. The ACURATE neo valve was associated with a lower risk of PPI at 30 days. No effect on 1-year mortality was determined for PPI at 30 days, while LBBB at 30 days was associated with reduced ejection fraction at 1 year.