A New Auto-RPA-Fluorescence Detection Platform for SARS-CoV-2

Lab Med. 2023 Mar 7;54(2):182-189. doi: 10.1093/labmed/lmac093.

Abstract

Objective: The outbreak of COVID-19 caused by SARS-CoV-2 has led to a serious worldwide pandemic. Quantitative reverse transcription-polymerase chain reaction (qRT-PCR)-based methods were recommended for routine detection of SARS-CoV-2 RNA. Because the reaction time and analytical sensitivity of qRT-PCR limits the diagnosis of SARS-CoV-2, development of a quick process of SARS-CoV-2 detection technology with high analytical sensitivity remains urgent.

Methods: We combined isothermal amplification and fluorescence detection technology to develop a new auto-recombinase polymerase amplification (RPA)-fluorescence platform that could be used in the diagnosis of SARS-CoV-2.

Results: By optimization of primers and probes, the RPA platform could detect SARS-CoV-2 nucleotides within 15 min. The limits of detection and specificity of the auto-RPA-fluorescence platform were 5 copies/µL and 100%, respectively. The accuracy of detection of the auto-RPA-fluorescence platform in the 16 positive samples was 100%.

Conclusion: The RPA platform is a potential technology for the diagnosis of SARS-CoV-2 infection.

Keywords: 2; CoV; RPA; SARS; auto; diagnosis; fluorescence; limit of detection; specificity.

MeSH terms

  • COVID-19* / diagnosis
  • Humans
  • RNA, Viral / genetics
  • Recombinases
  • SARS-CoV-2* / genetics
  • Sensitivity and Specificity