Assessment of aortic diameter in Marfan patients: intraindividual comparison of 3D-Dixon and 2D-SSFP magnetic resonance imaging

Eur Radiol. 2023 Mar;33(3):1687-1697. doi: 10.1007/s00330-022-09162-y. Epub 2022 Oct 21.

Abstract

Objectives: To compare the accuracy and precision of 3D-Dixon and 2D-SSFP MR-imaging for assessment of aortic diameter in Marfan patients.

Methods: This prospective single-center study investigated respiratory-gated 3D-Dixon and breath-hold 2D-SSFP non-contrast MR-imaging at 3 T in 47 Marfan patients (36.0 ± 13.2 years, 28♀,19♂). Two radiologists performed individual diameter measurements at five levels of the thoracic aorta and evaluated image quality on a four-grade scale (1 = poor, 4 = excellent) and artifacts (1 = severe, 4 = none). Aortic root diameters acquired by echocardiography served as a reference standard. Intraclass correlation coefficient, Bland-Altman analyses, F-test, t-test, and regression analyses were used to assess agreement between observers and methods.

Results: Greatest aortic diameters were observed at the level of the sinuses of Valsalva (SOV) for 3D-Dixon (38.2 ± 6.8 mm) and 2D-SSFP (38.3 ± 7.1 mm) (p = 0.53). Intra- and interobserver correlation of diameter measurements was excellent at all aortic levels for both 3D-Dixon (r = 0.94-0.99 and r = 0.94-0.98) and 2D-SSFP (r = 0.96-1.00 and r = 0.95-0.99). 3D-Dixon-derived and 2D-SSFP-derived diameter measurements at the level of the SOV revealed a strong correlation with echocardiographic measurements (r = 0.92, p < 0.001 and r = 0.93, p < 0.001, respectively). The estimated mean image quality at the level of SOV was higher for 2D-SSFP compared to that for 3D-Dixon (3.3 (95%-CI: 3.1-3.5) vs. 2.9 (95%-CI: 2.7-3.1)) (p < 0.001). Imaging artifacts were less at all aortic levels for 3D-Dixon compared to 2D-SSFP (3.4-3.8 vs. 2.8-3.1) (all p < 0.002).

Conclusion: Respiratory-gated 3D-Dixon and breath-hold 2D-SSFP MR-imaging provide accurate and precise aortic diameter measurements. We recommend 3D-Dixon imaging for monitoring of aortic diameter in Marfan patients due to fewer imaging artifacts and the possibility of orthogonal multiplanar reformations of the aortic root.

Key points: • Respiratory-gated 3D-Dixon and breath-hold 2D-SSFP imaging provide accurate and precise aortic diameter measurements in patients suffering from Marfan syndrome. • Imaging artifacts are stronger in 2D-SFFP imaging than in 3D-Dixon imaging. • We recommend 3D-Dixon imaging for monitoring of aortic diameter in Marfan patients due to fewer imaging artifacts and the possibility of orthogonal multiplanar reformations.

Keywords: Artifacts; Echocardiography; Magnetic resonance angiography; Marfan syndrome; Thoracic aorta.

MeSH terms

  • Contrast Media
  • Humans
  • Imaging, Three-Dimensional / methods
  • Magnetic Resonance Angiography / methods
  • Magnetic Resonance Imaging
  • Marfan Syndrome* / diagnostic imaging
  • Prospective Studies
  • Reproducibility of Results

Substances

  • Contrast Media