Purpose: Infant respiratory distress is a significant cause of mortality globally. Bubble continuous positive airway pressure (CPAP) is a simple and effective therapy, but sicker infants may require additional support such as non-invasive positive-pressure ventilation (NIPPV). We investigated the feasibility of a simple, low-cost, non-electric bubble NIPPV device.
Methods: In this cross-over feasibility study, seven newborns with moderate respiratory distress (Downes score ≥ 3), weight > 1500 g and gestational age > 32 weeks were randomized to 4 h of treatment with bubble CPAP (5-8 cm H2O) vs. bubble NIPPV (Phigh 8-10 cm H2O/Plow 5-8 cm H2O) followed by 4 h of the alternate treatment. Treatment order (CPAP vs. NIPPV) was randomized. Outcome measures included hourly vital signs, Downes score and O2 saturation. Adverse events including pneumothorax, nasal septal necrosis, necrotizing enterocolitis and death before discharge were also recorded.
Results: It took nurses 39 (7.3) s to assemble the bubble NIPPV device. Patients had similar vital signs and Downes scores on both treatments; median (IQR) values on bubble CPAP vs. bubble NIPPV were: heart rate 140 (134.5, 144), 140 (134.5, 144); respiratory rate 70 (56, 80), 65 (58, 82), Downes score 4 (3, 5.75), 4 (3, 5), O2 96 (94, 98), 97 (96, 98). All newborns survived to discharge and there were no adverse events. .
Conclusions: A simple, low-cost, non-electric method of providing NIPPV for newborns with respiratory distress is feasible in limited resource settings. Randomized-controlled trials comparing bubble CPAP and bubble NIPPV are justified.
Keywords: biomedical engineering; cost effective interventions; critical care; global health; newborn respiratory distress; non-invasive positive pressure ventilation (NIPPV).