Background: Despite its clear advantages over immunoassay-based testing, the measurement of serum thyroglobulin by mass spectrometry remains limited to a handful of institutions. Slow adoption by clinical laboratories could reflect limited accessibility to existing methods that have sensitivity comparable to modern immunoassays, as well as a lack of tools for calibration and assay harmonization.
Methods: We developed and validated a liquid chromatography-tandem mass spectrometry-based assay for the quantification of serum thyroglobulin. The protocol combined peptide immunoaffinity purification using a commercially available, well-characterized monoclonal antibody and mobile phase modification with dimethylsulfoxide (DMSO) for enhanced sensitivity. To facilitate harmonization with other laboratories, we developed a novel, serum-based 5-point distributable reference material (Husky Ref).
Results: The assay demonstrated a lower limit of quantification of 0.15 ng/mL (<20 %CV). Mobile phase DMSO increased signal intensity of the target peptide at least 3-fold, improving quantification at low concentrations. Calibration traceable to Husky Ref enabled harmonization between laboratories in an interlaboratory study.
Conclusions: Sensitive mass spectrometry-based thyroglobulin measurement can be achieved using a monoclonal antibody during peptide immunoaffinity purification and the addition of mobile phase DMSO. Laboratories interested in deploying this assay can utilize the provided standard operating procedure and freely-available Husky Ref reference material.
Keywords: Autoantibody; CPTAC, Clinical Proteomic Tumor Analysis Consortium; DMSO, dimethylsulfoxide; DTC, differentiated thyroid cancer; Differentiated thyroid cancer; IAE, immunoaffinity enrichment; LC-MS/MS, liquid chromatography-tandem mass spectrometry; Mass spectrometry; TLCK, tosyllysine chloromethyl ketone hydrochloride; Tg, thyroglobulin; Thyroglobulin.
© 2022 THE AUTHORS.