Experimental study of mosquito-inspired needle to minimize insertion force and tissue deformation

Proc Inst Mech Eng H. 2023 Jan;237(1):113-123. doi: 10.1177/09544119221137133. Epub 2022 Nov 27.

Abstract

The aim of this work is to propose a mosquito-inspired (bioinspired) design of a surgical needle that can decrease the insertion force and the tissue deformation, which are the main causes of target inaccuracy during percutaneous procedures. The bioinspired needle was developed by mimicking the geometrical shapes of mosquito proboscis. Needle prototypes were manufactured and tested to determine optimized needle shapes and geometries. Needle insertion tests on a tissue-mimicking polyvinylchloride (PVC) gel were then performed to emulate the mosquito-proboscis stinging dynamics by applying vibration and insertion velocity during the insertion. An insertion test setup equipped with a sensing system was constructed to measure the insertion force and to assess the deformation of the tissue. It was discovered that using the proposed bioinspired design, the needle insertion force was decreased by 60% and the tissue deformation was reduced by 48%. This finding is significant for improving needle-based medical procedures.

Keywords: Bioinspired surgical needle; dynamic insertion; insertion force; percutaneous procedures; tissue deformation.

MeSH terms

  • Mechanical Phenomena*
  • Needles*
  • Polyvinyl Chloride
  • Vibration

Substances

  • Polyvinyl Chloride