Interfractional dose accumulation for MR-guided liver SBRT: Variation among algorithms is highly patient- and fraction-dependent

Radiother Oncol. 2023 May:182:109448. doi: 10.1016/j.radonc.2022.109448. Epub 2022 Dec 22.

Abstract

Background and purpose: Daily plan adaptations could take the dose delivered in previous fractions into account. Due to high dose delivered per fraction, low number of fractions, steep dose gradients, and large interfractional organ deformations, this might be particularly important for liver SBRT. This study investigates inter-algorithm variation of interfractional dose accumulation for MR-guided liver SBRT.

Materials and methods: We assessed 27 consecutive MR-guided liver SBRT treatments of 67.5 Gy in three (n = 15) or 50 Gy in five fractions (n = 12), both prescribed to the GTV. We calculated fraction doses on daily patient anatomy, warped these doses to the simulation MRI using seven different algorithms, and accumulated the warped doses. Thus, we obtained differences in planned doses and warped or accumulated doses for each algorithm. This enabled us to calculate the inter-algorithm variations in warped doses per fraction and in accumulated doses per treatment course.

Results: The four intensity-based algorithms were more consistent with planned PTV dose than affine or contour-based algorithms. The mean (range) variation of the dose difference for PTV D95% due to dose warping by these intensity-based algorithms was 10.4 percentage points (0.3 to 43.7) between fractions and 8.6 (0.3 to 24.9) between accumulated treatment doses. As seen by these ranges, the variation was very dependent on the patient and the fraction being analyzed. Nevertheless, no correlations between patient or plan characteristics on the one hand and inter-algorithm dose warping variation on the other hand was found.

Conclusion: Inter-algorithm dose accumulation variation is highly patient- and fraction-dependent for MR-guided liver SBRT. We advise against trusting a single algorithm for dose accumulation in liver SBRT.

Keywords: Deformable image registration; Dose accumulation; Liver metastases; MR-guided radiotherapy; Radiotherapy; SABR.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Humans
  • Liver / diagnostic imaging
  • Radiosurgery*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy, Intensity-Modulated*