Objective.To examine the influence of the endotracheal tube (ETT) on respiratory reactance (Xrs) measured with the forced oscillation technique (FOT) and develop a correction method for it.Approach.In a bench study, the reactance of ETTs (Xtube) with different dimensions was measured on a breathing test lung in various respiratory settings.Main results.Xtubecan be accurately predicted by a fitted formula, with an R2of 0.97, with negligible effects due to changes in respiratory pattern and lung volume.Significance.The developed formula offers the ability to measure ETT-independent Xrsvalues of patients, improving the potential of FOT for lung function testing in mechanically ventilated newborns.
Keywords: infants; mechanical ventilation; oscillometry; respiratory mechanics.
Creative Commons Attribution license.