Correction of the urinary testosterone to epitestosterone ratio measurement in antidoping analyses by chromatographic and mass spectrometric techniques

Drug Test Anal. 2023 Nov-Dec;15(11-12):1382-1391. doi: 10.1002/dta.3456. Epub 2023 Feb 15.

Abstract

The ratio of testosterone (T) to epitestosterone (E) determined in urine samples is the main biomarker used to prove T and precursors abused by athletes. Analytically, the correction of the ratio is required by the World Anti-Doping Agency. This work describes a series of experiments aimed to study when it is appropriate to correct the T/E ratio value, using different mass spectrometric techniques since only reports on GC-MS exist. Analyses using external calibrators, controls, and routine samples were performed by three different techniques GC-MS, GC-MSn , and LC-MSn . A statistical comparison of the T/E was performed after the application of two corrections previously published: Isotopic contribution peak area correction (corr_1) and use of a verified internal deuterated internal standard (TD3/ED3) correction (corr_2) and the ratio based on T and E concentrations. The use of external calibration samples introduces biases that influence not only the T and E concentrations but also the T/E ratio, even when both methods of correction are applied. The correction after applying corr_1 method barely contributed to the accuracy of the T/E calculation. Nevertheless, the application of the corr_2 method increased the accuracy between 5% and 6% when comparing theoretical and experimental T/E values. Finally, the best results were obtained by the ratio calculated directly from the estimated concentrations of T and E. Attention must be paid when the T/E is measured by LC-MSn since different acquisition modes produced significantly different results, even in MS/MS when the same transition is used for T and E.

Keywords: T/E ratio; TD-EAAS; gas chromatography; isotopic correction; liquid chromatography.