Preanalytical Errors in a Hematology Laboratory: An Experience from a Tertiary Care Center

Diagnostics (Basel). 2023 Feb 6;13(4):591. doi: 10.3390/diagnostics13040591.

Abstract

Background: Laboratory errors arise at any stage of testing. Detecting these inaccuracies before results are revealed might delay diagnosis and treatment, causing patient distress. Here, we studied the preanalytical errors in a hematology laboratory.

Methods: This one-year retrospective analysis was conducted at the laboratory of a tertiary care hospital and included information on blood samples that were taken for hematology tests from both outpatients and inpatients. Laboratory records included sample collection and rejection information. The type and frequency of preanalytical errors were expressed as a proportion of total errors and sample number. Microsoft Excel was utilized to enter data. The results were presented in the form of frequency tables.

Results: This research included 67,892 hematology samples. For preanalytical errors, 886 samples (1.3%) were discarded. The most common preanalytical error was insufficient sample (54.17%), and the least common was an empty/damaged tube (0.4%). Erroneous samples in the emergency department were mostly insufficient and clotted, whereas pediatric sample errors were caused by insufficient and diluted samples.

Conclusion: Inadequate samples and clotted samples account for the vast majority of preanalytical factors. Insufficiency and dilutional errors were most frequent from pediatric patients. Adherence to best laboratory practices can drastically cut down on preanalytical errors.

Keywords: hematology; laboratory; preanalytical error; quality assurance.