Identifying Parathyroids in Pediatric Thyroid/Parathyroid Surgery by Near Infrared Autofluorescence

Laryngoscope. 2023 Nov;133(11):3208-3215. doi: 10.1002/lary.30633. Epub 2023 Mar 3.

Abstract

Objectives: Compared to adult patients undergoing thyroid surgery, pediatric patients have higher rates of hypoparathyroidism often related to parathyroid gland (PG) inadvertent injury or devascularization. Previous studies have shown that near-infrared-autofluorescence (NIRAF) can be reliably used intraoperatively for label-free parathyroid identification, but all prior studies have been performed in adult patients. In this study, we assess the utility and accuracy of NIRAF with a fiber-optic probe-based system to identify PGs in pediatric patients undergoing thyroidectomy or parathyroidectomy.

Methods: All pediatric patients (under 18 years of age) undergoing thyroidectomy or parathyroidectomy were enrolled in this IRB-approved study. The surgeon's visual assessment of tissues was first noted and the surgeon's confidence level in the tissue identified was recorded. A fiber-optic probe was then used to illuminate tissues-of-interest with a wavelength of 785 nm and resulting NIRAF intensities from these tissues were measured while the surgeon was blinded to results.

Results: NIRAF intensities were measured intraoperatively in 19 pediatric patients. Normalized NIRAF intensities for PGs (3.63 ± 2.47) were significantly higher than that of thyroid (0.99 ± 0.36, p < 0.001) and other surrounding soft tissues (0.86 ± 0.40, p < 0.001). Based on the PG identification ratio threshold of 1.2, NIRAF yielded a detection rate of 95.8% (46/48 pediatric PGs).

Conclusion: Our findings indicate that NIRAF detection can potentially be a valuable and non-invasive technique to identify PGs during neck operations in the pediatric population. To our knowledge, this is the first study in children to assess the accuracy of probe-based NIRAF detection for intraoperative parathyroid identification.

Level of evidence: Level 4 Laryngoscope, 133:3208-3215, 2023.

Keywords: adolescent; autofluorescence; infrared; parathyroid; pediatric; thyroid.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Adolescent
  • Adult
  • Child
  • Humans
  • Optical Imaging / methods
  • Parathyroid Glands* / diagnostic imaging
  • Parathyroid Glands* / injuries
  • Parathyroid Glands* / surgery
  • Parathyroidectomy / methods
  • Thyroid Gland* / surgery
  • Thyroidectomy / adverse effects