The PETcoil project: PET performance evaluation of two detector modules for a second generation RF-penetrable TOF-PET brain dedicated insert for simultaneous PET/MRI
- PMID: 36913739
- DOI: 10.1088/1361-6560/acc3f2
The PETcoil project: PET performance evaluation of two detector modules for a second generation RF-penetrable TOF-PET brain dedicated insert for simultaneous PET/MRI
Abstract
Objective. We are developing a portable, 'RF-penetrable', brain-dedicated time of flight (TOF)-PET insert (PETcoil) for simultaneous PET/MRI.Approach. In this paper, we evaluate the PET performance of two fully assembled detector modules for this insert design outside the MR room.Main results. The global coincidence time resolution, global 511 keV energy resolution, coincidence count rate, and detector temperature achieved over 2 h data collection were 242.2 ± 0.4 ps full width at half maximum (FWHM), 11.19% ± 0.02% FWHM, 22.0 ± 0.1 kcps, and 23.5 °C ± 0.3 °C, respectively. The intrinsic spatial resolutions in the axial and transaxial directions were 2.74 ± 0.01 mm FWHM and 2.88 ± 0.03 mm FWHM, respectively.Significance. These results demonstrate excellent TOF capability and the performance and stability necessary for scaling up to a full ring comprising 16 detector modules.
Keywords: MR-compatible; PET insert; PETcoil; RF-penetrable; TOF; brain imaging; simultaneous PET/MRI.
© 2023 Institute of Physics and Engineering in Medicine.
Similar articles
-
Study of compatibility between a 3T MR system and detector modules for a second-generation RF-penetrable TOF-PET insert for simultaneous PET/MRI.Med Phys. 2023 Jun;50(6):3389-3400. doi: 10.1002/mp.16354. Epub 2023 Apr 7. Med Phys. 2023. PMID: 36912373
-
PETcoil: first results from a second-generation RF-penetrable TOF-PET brain insert for simultaneous PET/MRI.Phys Med Biol. 2024 Sep 9;69(18). doi: 10.1088/1361-6560/ad7221. Phys Med Biol. 2024. PMID: 39168156
-
Performance study of a radio-frequency field-penetrable PET insert for simultaneous PET/MRI.IEEE Trans Radiat Plasma Med Sci. 2018 Sep;2(5):422-431. doi: 10.1109/TRPMS.2018.2852686. Epub 2018 Jul 3. IEEE Trans Radiat Plasma Med Sci. 2018. PMID: 30911706 Free PMC article.
-
Performance evaluation of a high-resolution brain PET scanner using four-layer MPPC DOI detectors.Phys Med Biol. 2017 Aug 18;62(17):7148-7166. doi: 10.1088/1361-6560/aa82e8. Phys Med Biol. 2017. PMID: 28753133
-
Sub-3 mm, near-200 ps TOF/DOI-PET imaging with monolithic scintillator detectors in a 70 cm diameter tomographic setup.Phys Med Biol. 2018 Jul 27;63(15):155006. doi: 10.1088/1361-6560/aad2a6. Phys Med Biol. 2018. PMID: 29995639
Cited by
-
New Horizons in Brain PET Instrumentation.PET Clin. 2024 Jan;19(1):25-36. doi: 10.1016/j.cpet.2023.08.001. Epub 2023 Oct 6. PET Clin. 2024. PMID: 37806894 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical