Clinical Feasibility of Using Single-isocentre Non-coplanar Volumetric Modulated Arc Therapy Combined with Non-coplanar Cone Beam Computed Tomography in Hypofractionated Stereotactic Radiotherapy for Five or Fewer Multiple Intracranial Metastases

Clin Oncol (R Coll Radiol). 2023 Jun;35(6):408-416. doi: 10.1016/j.clon.2023.03.011. Epub 2023 Mar 22.

Abstract

Aims: To evaluate the clinical feasibility of single-isocentre non-coplanar volumetric modulated arc therapy (NC-VMAT) with non-coplanar cone beam computed tomography (NC-CBCT) in hypofractionated stereotactic radiotherapy (HSRT) for five or fewer multiple brain metastases.

Materials and methods: Ten patients with multiple brain metastases who underwent single-isocentre NC-VMAT HSRT with limited couch rotations (within ±45°) and NC-CBCT with a limited scanning range (150-200°) were included in the current analysis. Conventional single-isocentre coplanar VMAT (C-VMAT) plans were generated and compared with NC-VMAT plans. The intracranial response and toxicities of single-isocentre NC-VMAT HSRT were also evaluated.

Results: Compared with C-VMAT, NC-VMAT generated better target conformity (P < 0.05), a lower gradient index (P < 0.05) and better normal brain tissue sparing, especially for volume ≥12 Gy, with a median reduction of 12.65 cm3. For 45° couch rotation, NC-CBCT produced sufficient image quality to differentiate bony anatomy, even with a 150° scanning range, which could be successfully used for patient set-up correction. After NC-CBCT, 57.1% of the measured non-coplanar set-up errors exceeded the threshold value. The median gamma passing rate of NC-VMAT was higher than that of C-VMAT plans (P < 0.05). The non-coplanar beam of NC-VMAT with NC-CBCT corrections exhibited superior gamma passing rate to that without NC-CBCT corrections. The intracranial objective response rate and disease control rate for all patients were 80% (8/10) and 100% (10/10), respectively, and the most common toxicities were headache (20%) and dizziness (20%).

Conclusion: NC-VMAT with limited couch rotation (within ±45°) combined with NC-CBCT with a limited scanning range (150-200°) markedly improves the plan quality and set-up accuracy in single-isocentre multiple-target HSRT.

Keywords: Hypofractionated stereotactic radiotherapy; multiple brain metastases; non-coplanar cone beam computed tomography; plan quality; set-up accuracy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain Neoplasms* / diagnostic imaging
  • Brain Neoplasms* / radiotherapy
  • Brain Neoplasms* / secondary
  • Cone-Beam Computed Tomography
  • Feasibility Studies
  • Humans
  • Radiosurgery* / methods
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods
  • Radiotherapy, Intensity-Modulated* / methods