Coronary Stent Off-Wire Dislodgement: Case Report of a Rare Complication

Tex Heart Inst J. 2023 Mar 1;50(2):e207505. doi: 10.14503/THIJ-20-7505.

Abstract

Although several techniques have been reported for managing an on-wire dislodged stent in the coronary artery, very few reports have focused on the much rarer complication of an off-wire dislodged stent. In a 73-year-old man who experienced an off-wire dislodged coronary stent, the proximal elongated segment was lodged in the left main coronary artery, and the distal segment was floating in the aorta like a wind sock. After a failed attempt at retrieval using a gooseneck microsnare, the dislodged stent was successfully removed using a 3-loop vascular snare via the left radial artery. There was no obvious vascular injury. This novel technique for removing a partially floating dislodged stent was successful after conventional retrieval techniques failed.

Keywords: Aorta; complication, intraoperative; drug eluting stent; endovascular procedure; percutaneous coronary intervention.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary* / adverse effects
  • Angioplasty, Balloon, Coronary* / methods
  • Coronary Angiography
  • Coronary Vessels / diagnostic imaging
  • Coronary Vessels / surgery
  • Humans
  • Male
  • Stents

Grants and funding

Funding/Support: None declared