Diagnostic Performance of BD Phoenix CPO Detect Panels for Detection and Classification of Carbapenemase-Producing Gram-Negative Bacteria

Microbiol Spectr. 2023 Jun 15;11(3):e0089723. doi: 10.1128/spectrum.00897-23. Epub 2023 May 10.

Abstract

BD Phoenix CPO Detect panels can identify and classify carbapenemase-producing organisms (CPOs) simultaneously with antimicrobial susceptibility testing (AST) for Gram-negative bacteria. Detection and classification of carbapenemase producers were performed using the BD Phoenix CPO Detect panels NMIC/ID-441 for Enterobacterales, NMIC/ID-442 for nonfermenting bacteria, and NMIC-440 for both. The results were compared with those obtained using comparator methods. A total of 133 strains (32 Klebsiella pneumoniae, 37 Enterobacter cloacae complex, 33 Pseudomonas aeruginosa, and 31 Acinetobacter baumannii complex strains), including 60 carbapenemase producers (54 imipenemases [IMPs] and 6 OXA type), were analyzed. Using panels NMIC-440 and NMIC/ID-441 or NMIC/ID-442, all 54 IMP producers were accurately identified as CPOs (positive percent agreement [PPA], 100.0%; 54/54). Among the 54 IMP producers identified as CPOs using panels NMIC-440 and NMIC/ID-441, 12 and 14 Enterobacterales were not resistant to carbapenem, respectively. Among all 54 IMP producers, 48 (88.9%; 48/54) were correctly classified as Ambler class B using panel NMIC-440. Using panels NMIC-440 and NMIC/ID-442, all four OXA-23-like carbapenemase-producing A. baumannii complex strains (100.0%, 4/4) were correctly identified as CPOs, and three (75.0%, 3/4) were precisely classified as class D using panel NMIC-440. Both carbapenemase producers harboring the blaISAba1-OXA-51-like gene were incorrectly identified as non-CPOs using panels NMIC-440 and NMIC/ID-442. For detecting carbapenemase producers, the overall PPA and negative percent agreement (NPA) between panel NMIC-440 and the comparator methods were 96.7% (58/60) and 71.2% (52/73), respectively, and the PPA and NPA between panels NMIC/ID-441 or NMIC/ID-442 and the comparator methods were 96.7% (58/60) and 74.0% (54/73), respectively. BD Phoenix CPO Detect panels can successfully screen carbapenemase producers, particularly IMP producers, regardless of the presence of carbapenem resistance and can be beneficial in routine AST workflows. IMPORTANCE Simple and efficient screening methods of detecting carbapenemase producers are required. BD Phoenix CPO Detect panels effectively screened carbapenemase producers, particularly IMP producers, with a high overall PPA. As the panels enable automatic screening for carbapenemase producers simultaneously with AST, the workflow from AST to confirmatory testing for carbapenemase production can be shortened. In addition, because carbapenem resistance varies among carbapenemase producers, the BD Phoenix CPO Detect panels, which can screen carbapenemase producers regardless of carbapenem susceptibility, can contribute to the accurate detection of carbapenemase producers. Our results report that these panels can help streamline the AST workflow before confirmatory testing for carbapenemase production in routine microbiological tests.

Keywords: Ambler classification; automated system; carbapenem resistance; carbapenemase; rapid detection.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Anti-Bacterial Agents / pharmacology
  • Bacterial Proteins / genetics
  • Carbapenem-Resistant Enterobacteriaceae* / classification
  • Carbapenem-Resistant Enterobacteriaceae* / drug effects
  • Carbapenem-Resistant Enterobacteriaceae* / isolation & purification
  • Carbapenems / pharmacology
  • Microbial Sensitivity Tests
  • Microbiological Techniques* / methods
  • Microbiological Techniques* / standards
  • Reproducibility of Results
  • Species Specificity

Substances

  • Anti-Bacterial Agents
  • Bacterial Proteins
  • carbapenemase
  • Carbapenems