Comparison of first- and second-generation leadless pacemakers in patients with sinus rhythm and complete atrioventricular block

J Cardiovasc Electrophysiol. 2023 Aug;34(8):1730-1737. doi: 10.1111/jce.15981. Epub 2023 Jun 24.

Abstract

Introduction: The efficacy and safety of leadless cardiac pacemakers (LPMs) as an alternative to conventional transvenous cardiac pacing have been largely reported. The first generation of the MicraTM transcatheter pacing system (VR; Medtronic) was able to provide single-chamber VVI(R) pacing mode only, with a potential risk of pacemaker syndrome in sinus rhythm patients. A second-generation system (AV) now provides atrioventricular synchrony through atrial mechanical (Am) sensing capability (VDD mode).

Objective: We sought to compare VR and AV systems in sinus rhythm patients with chronic ventricular pacing (Vp) for complete atrioventricular block.

Methods: All consecutive patients implanted with an LPM in our department for complete atrioventricular block were retrospectively screened. Patients with atrial fibrillation, sinus dysfunction, or Vp burden <20% at 1 month postimplantation were excluded. Patients were systematically followed with a visit at 1 month, and then at least once a year.

Results: A total of 93 patients-45 VR (2015-2020) and 48 AV (2020-2021)-were included. VR and AV patients had similar baseline characteristics, except for VR patients being older (80 ± 8 vs. 77 ± 9 years, p = 0.049). The mean Vp burden was 77% in the VR and 82% in the AV group (p = 0.38). In AV patients, the median AV synchronous beats rate was 78%, with 65% having a >66% rate. An E/A ratio <1.2 as measured on echocardiography was the only independent predictor of accurate atrial mechanical tracking (p = 0.01). One-year survival rate was similar in both groups. Five patients in the VR and 0 in the AV group eventually developed pacemaker syndrome within 1 year post-implantation (p = 0.02).

Conclusion: In sinus rhythm patients with chronic Vp for complete atrioventricular block implanted with an LPM, the atrial mechanical sensing algorithm allowed significant atrioventricular synchrony in most patients and was associated with no occurrence of-otherwise rare-pacemaker syndrome.

Keywords: atrioventricular synchrony; leadless cardiac pacing; pacemaker syndrome.

MeSH terms

  • Atrial Fibrillation* / diagnosis
  • Atrial Fibrillation* / epidemiology
  • Atrial Fibrillation* / therapy
  • Atrioventricular Block* / diagnosis
  • Atrioventricular Block* / therapy
  • Cardiac Pacing, Artificial / adverse effects
  • Heart Atria
  • Humans
  • Pacemaker, Artificial*
  • Postoperative Complications
  • Retrospective Studies