Temporal, Kinematic and Kinetic Variables Derived from a Wearable 3D Inertial Sensor to Estimate Muscle Power during the 5 Sit to Stand Test in Older Individuals: A Validation Study

Sensors (Basel). 2023 May 16;23(10):4802. doi: 10.3390/s23104802.

Abstract

The 5-Sit-to-stand test (5STS) is widely used to estimate lower limb muscle power (MP). An Inertial Measurement Unit (IMU) could be used to obtain objective, accurate and automatic measures of lower limb MP. In 62 older adults (30 F, 66 ± 6 years) we compared (paired t-test, Pearson's correlation coefficient, and Bland-Altman analysis) IMU-based estimates of total trial time (totT), mean concentric time (McT), velocity (McV), force (McF), and MP against laboratory equipment (Lab). While significantly different, Lab vs. IMU measures of totT (8.97 ± 2.44 vs. 8.86 ± 2.45 s, p = 0.003), McV (0.35 ± 0.09 vs. 0.27 ± 0.10 m∙s-1, p < 0.001), McF (673.13 ± 146.43 vs. 653.41 ± 144.58 N, p < 0.001) and MP (233.00 ± 70.83 vs. 174.84 ± 71.16 W, p < 0.001) had a very large to extremely large correlation (r = 0.99, r = 0.93, and r = 0.97 r = 0.76 and r = 0.79, respectively, for totT, McT, McF, McV and MP). Bland-Altman analysis showed a small, significant bias and good precision for all the variables, but McT. A sensor-based 5STS evaluation appears to be a promising objective and digitalized measure of MP. This approach could offer a practical alternative to the gold standard methods used to measure MP.

Keywords: IMU; ageing; functional screening; motion analysis; sensors; sit-to-stand; validity.

MeSH terms

  • Aged
  • Biomechanical Phenomena
  • Female
  • Humans
  • Kinetics
  • Lower Extremity*
  • Male
  • Middle Aged
  • Muscles
  • Wearable Electronic Devices*