Multicenter matched-pair study comparing BACT/ALERT® MP reagent systems for the recovery of mycobacteria from specimens other than blood

Diagn Microbiol Infect Dis. 2023 Oct;107(2):115959. doi: 10.1016/j.diagmicrobio.2023.115959. Epub 2023 Apr 20.

Abstract

The BACT/ALERT® MP Reagent System is a broth culture medium for optimal detection and recovery of mycobacteria from clinical samples. The MP formulation was recently modified to improve detection and recovery times. A multicenter prospective matched pair study design was conducted to validate the performance of improved MP (MP-I) versus current MP (MP-C) bottles utilizing nonsterile and normally sterile samples, except blood, from patients suspected of having mycobacterial infections. A total of 1488 clinical samples were collected to obtain 212 mycobacteria samples by either or both MP culture bottles. MP-I and MP-C sensitivities were 86.6% and 81.4%, respectively, but the difference was not significant (P = 0.163) while specificities were 96.8% and 93.8%, respectively, and that difference was significant (P = 0.002). Overall recovery was 94.34% for MP-I and 88.68% for MP-C (recovery was 100% for both bottles with 52 seeded samples). Overall performance of MP-I was better than MP-C for sensitivity, specificity, and recovery.

Keywords: Culture; Detection; Mycobacteria.

Publication types

  • Multicenter Study

MeSH terms

  • Culture Media
  • Humans
  • Mycobacterium Infections* / microbiology
  • Mycobacterium*
  • Prospective Studies
  • Reagent Kits, Diagnostic

Substances

  • Culture Media
  • Reagent Kits, Diagnostic