Time-resolved cryo-EM using a combination of droplet microfluidics with on-demand jetting

Nat Methods. 2023 Sep;20(9):1400-1408. doi: 10.1038/s41592-023-01967-z. Epub 2023 Aug 17.

Abstract

Single-particle cryogenic electron microscopy (cryo-EM) allows reconstruction of high-resolution structures of proteins in different conformations. Protein function often involves transient functional conformations, which can be resolved using time-resolved cryo-EM (trEM). In trEM, reactions are arrested after a defined delay time by rapid vitrification of protein solution on the EM grid. Despite the increasing interest in trEM among the cryo-EM community, making trEM samples with a time resolution below 100 ms remains challenging. Here we report the design and the realization of a time-resolved cryo-plunger that combines a droplet-based microfluidic mixer with a laser-induced generator of microjets that allows rapid reaction initiation and plunge-freezing of cryo-EM grids. Using this approach, a time resolution of 5 ms was achieved and the protein density map was reconstructed to a resolution of 2.1 Å. trEM experiments on GroEL:GroES chaperonin complex resolved the kinetics of the complex formation and visualized putative short-lived conformations of GroEL-ATP complex.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cognition*
  • Computer Systems
  • Cryoelectron Microscopy
  • Electrons
  • Microfluidics*