Different Home Blood Pressure Thresholds to Predict Perfect 24-Hour Ambulatory Blood Pressure Control in Treated Hypertension Based on an "All-in-One" Device

Hypertension. 2023 Nov;80(11):2464-2472. doi: 10.1161/HYPERTENSIONAHA.123.21578. Epub 2023 Sep 6.

Abstract

Background: Home blood pressure (BP) is an important component of digital strategies for hypertension management. However, no studies have used the same device to investigate 24-hour BP control status in relation to different home BP control thresholds.

Methods: Participants in the general practitioner-based, multicenter HI-JAMP study (Home-Activity Information and Communication Technology-Based Japan Ambulatory Blood Pressure Monitoring Prospective) underwent office BP measurement, then 24-hour ambulatory BP monitoring, then home BP monitoring for 5 days. A validated all-in-one BP monitoring device was used to measure office, home, and ambulatory BP. Baseline data were used to investigate ambulatory BP control status in individuals with well-controlled home BP based on the different guideline thresholds (125/75 mm Hg, 130/80 mm Hg, and 135/85 mm Hg).

Results: Data from 2269 patients were analyzed. For individuals with well-controlled home BP <135/85 mm Hg (59.5% of the total population), the prevalence of uncontrolled 24-hour (≥130/80 mm Hg), daytime (≥135/85 mm Hg), and nighttime ambulatory BP (≥120/70 mm Hg) was 19.9%, 18.5%, and 33.6%, respectively. Corresponding prevalence rates in the 42.7% of participants with well-controlled home BP <130/80 mm Hg were 13.4%, 12.9%, and 26.0%, and when well-controlled home BP was strictly defined as <125/75 mm Hg (23.9% of the population), prevalence of rates of uncontrolled 24-hour, daytime, and nighttime ambulatory BP were 7.0%, 9.0%, and 15.3%, respectively.

Conclusions: Home BP control status defined using different thresholds could predict 24-hour ambulatory BP control status in treated hypertension. One-third of individuals still had uncontrolled nocturnal hypertension when home BP was controlled to <135/85 mm Hg, but ambulatory BP was quite well controlled when home BP was <125/75 mm Hg.

Keywords: ambulatory blood pressure monitoring; blood pressure; blood pressure monitors; hypertension; masked hypertension.