Deep learning-based iodine contrast-augmenting algorithm for low-contrast-dose liver CT to assess hypovascular hepatic metastasis

Abdom Radiol (NY). 2023 Nov;48(11):3430-3440. doi: 10.1007/s00261-023-04039-0. Epub 2023 Sep 14.

Abstract

Purpose: To investigate the image quality and diagnostic performance of low-contrast-dose liver CT using a deep learning-based iodine contrast-augmenting algorithm (DLICA) for hypovascular hepatic metastases.

Methods: This retrospective study included 128 patients who underwent contrast-enhanced dual-energy CT for hepatic metastasis surveillance between July 2019 and June 2022 using a 30% reduced iodine contrast dose in the portal phase. Three image types were reconstructed: 50-keV virtual monoenergetic images (50-keV VMI); linearly blended images simulating 120-kVp images (120-kVp); and post-processed 120-kVp images using DLICA (DLICA 120-kVp). Three reviewers evaluated lesion conspicuity, image contrast, and subjective image noise. We also measured image noise, contrast-to-noise ratios (CNRs), and signal-to-noise ratios (SNRs). The diagnostic performance for hepatic metastases was evaluated using a jackknife alternative free-response receiver operating characteristic method with the consensus of two independent radiologists as the reference standard.

Results: DLICA 120-kVp demonstrated significantly higher CNR of lesions to liver (5.7 ± 3.1 vs. 3.8 ± 2.1 vs. 3.8 ± 2.1) and higher SNR compared with 50-keV VMI and 120-kVp (p < 0.001 for all). DLICA 120-kVp had significantly lower image noise than 50-kVp VMI for all regions (p < 0.001 for all). DLICA 120-kVp also exhibited superior lesion conspicuity (4.0 [3.3-4.3] vs. 3.7 [3.0-4.0] vs. 3.7 [3.0-4.0]), higher image contrast, and lower subjective image noise compared with 50-keV VMI and 120-kVp (p < 0.001 for all). Although there was no significant difference in the figure of merit for lesion diagnosis among the three methods (p = 0.11), DLICA 120-kVp had a significantly higher figure of merit for lesions with a diameter < 20 mm than 50-keV VMI (0.677 vs. 0.648, p = 0.007). On a per-lesion basis, DLICA 120-kVp also demonstrated higher sensitivity than the 50-keV VMI (81.2% vs. 72.9%, p < 0.001). The specificities per lesion were not significantly different among the three algorithms (p = 0.15).

Conclusion: DLICA at 120-kVp provided superior lesion conspicuity and image quality and similar diagnostic performance for hypovascular hepatic metastases compared with 50-keV VMI.

Keywords: Contrast media; Deep learning; Liver; Neoplasm metastasis; X-ray computed tomography.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Deep Learning*
  • Humans
  • Iodine*
  • Liver Neoplasms* / diagnostic imaging
  • Liver Neoplasms* / secondary
  • Radiographic Image Interpretation, Computer-Assisted / methods
  • Radiography, Dual-Energy Scanned Projection* / methods
  • Retrospective Studies
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed / methods

Substances

  • Iodine
  • Iodine-120