Diagnostic Performance and Safety Profile of Robotic-assisted Bronchoscopy: A Systematic Review and Meta-Analysis

Ann Am Thorac Soc. 2023 Dec;20(12):1801-1812. doi: 10.1513/AnnalsATS.202301-075OC.

Abstract

Rationale: Conventional electromagnetic navigation bronchoscopy and other guided bronchoscopic modalities have a very desirable safety profile, but their diagnostic yield is only 60-70% for pulmonary lesions. Recently, robotic-assisted bronchoscopy (RAB) platforms have been introduced to improve the diagnostic performance of bronchoscopic modalities. Objectives: To determine the diagnostic performance and safety profile of RAB (using shape-sensing and electromagnetic navigation-based platforms) by performing a systematic review and meta-analysis. Methods: The PubMed, Embase, and Google Scholar databases were searched to find studies that reported on the diagnostic performance and/or the safety profile of one of the RAB systems. The quality of included studies was assessed using the Quality Assessment of Diagnostic Accuracy Studies 2 tool. Meta-analysis was performed using MedCalc version 20.118. Pooled diagnostic yield was calculated using a Freeman-Tukey transformation. We planned to use a random-effects model if the I2 index was >40%. Results: Twenty-five studies were included: 20 including diagnostic and safety analyses and 5 including only safety analyses. The pooled diagnostic yield of RAB (20 studies, 1,779 lesions) was 84.3% (95% confidence interval, 81.1-87.2%). The I2 index was 65.6%. On the basis of our subgroup analyses, the heterogeneity was likely driven by differences in study designs (prospective vs. retrospective) and procedural protocols (such as different RAB systems). Lesion size > 2 cm, the presence of a computed tomography bronchus sign, and concentric radial endobronchial ultrasound view were associated with a statistically significant increase in the odds of diagnosis with RAB. The overall rates of pneumothorax, need for tube thoracostomy, and significant hemorrhage were 2.3%, 1.2%, and 0.5%, respectively. Conclusions: RAB systems have significantly increased the diagnostic yield of navigational bronchoscopy compared with conventional systems such as electromagnetic navigation bronchoscopy, but well-designed prospective studies are needed to better understand the impact of various factors, such as the use of three-dimensional imaging modalities, cryobiopsy, and specific ventilatory protocols, on the diagnostic yield of RAB.

Keywords: electromagnetic navigation robotic-assisted bronchoscopy; lung nodule; navigational bronchoscopy; robotic bronchoscopy; shape-sensing robotic-assisted bronchoscopy.

Publication types

  • Meta-Analysis
  • Systematic Review

MeSH terms

  • Bronchoscopy / adverse effects
  • Bronchoscopy / methods
  • Humans
  • Lung Neoplasms* / diagnosis
  • Prospective Studies
  • Retrospective Studies
  • Robotic Surgical Procedures* / adverse effects