Background: The NHS has the target of reducing its carbon emission by 80% by 2032. Part of its strategy is using pharmaceuticals with a less harmful impact on the environment. Nitrous oxide is currently used widely within the NHS. Nitrous oxide, if released into the atmosphere, has a significant environmental impact. Methoxyflurane, delivered through the Penthrox 'green whistle' device, is a short-acting analgesic and is thought to have a smaller environmental impact compared with nitrous oxide.
Methods: Life cycle impact assessment (LCIA) of all products and processes involved in the manufacture and use of Penthrox, using data from the manufacturer, online sources and LCIA inventory Ecoinvent. These data were analysed in OpenLCA. Impact data were compared with existing data on nitrous oxide and morphine sulfate.
Results: This LCIA found that Penthrox has a climate change effect of 0.84 kg carbon dioxide equivalent (CO2e). Raw materials and the production process contributed to majority of the impact of Penthrox across all categories with raw materials accounting for 34.40% of the total climate change impact. Penthrox has a climate change impact of 117.7 times less CO2e compared with Entonox. 7 mg of 100 mg/100 mL of intravenous morphine sulfate had a climate change effect of 0.01 kg CO2e.
Conclusions: This LCIA has shown that the overall 'cradle-to-grave' environmental impact of Penthrox device is better than nitrous oxide when looking specifically at climate change impact. The climate change impact for an equivalent dose of intravenous morphine was even lower. Switching to the use of inhaled methoxyflurane instead of using nitrous oxide in certain clinical situations could help the NHS to reach its carbon emission reduction target.
Keywords: analgesia; emergency departments; pain management.
© Author(s) (or their employer(s)) 2024. No commercial re-use. See rights and permissions. Published by BMJ.