Surface Modification of Silica with β-Alanine Derivatives for Unique Applications in Liquid Chromatography

ACS Appl Mater Interfaces. 2023 Nov 22;15(46):54176-54184. doi: 10.1021/acsami.3c11932. Epub 2023 Nov 10.

Abstract

Column purchasing cost is an important issue for an analyst to analyze complex sample matrices. Here, we report the development of an amino acid (β-alanine)-derived stationary phase (Sil-Ala-C12) with strategic and effective interaction sites (amide and urea as embedded polar groups with C12 alkyl chain) able to separate various kinds of analytes. Owing to the balanced hydrophobicity and hydrophilicity of the phase, it showed exceptional separation abilities in both reversed-phase high-performance liquid chromatography (RP-HPLC) as a hydrophobic phase and hydrophilic interaction chromatography (HILIC) as a hydrophilic phase. Remarkably, the baseline separation was achieved for the challenging β- and γ-isomers of tocopherol. Usually, three columns such as pentafluorophenyl or C30, C18, and sulfobetaine HILIC are required for the analysis of vitamin E, capsaicinoids, and vitamin C in chili peppers (Capsicum spp.), respectively. However, only Sil-Ala-C12 was able to separate these analytes. A single column can serve 3-4 purposes, which suggests that Sil-Ala-C12 had the potential to reduce column purchasing costs.

Keywords: HILIC; chili pepper; colum purchasing cost; polar embedded group; tocopherol.

MeSH terms

  • Amides / chemistry
  • Chromatography, Liquid / methods
  • Chromatography, Reverse-Phase* / methods
  • Hydrophobic and Hydrophilic Interactions
  • Silicon Dioxide* / chemistry
  • beta-Alanine

Substances

  • Silicon Dioxide
  • Amides
  • beta-Alanine