m-PIMA™ HIV1/2 VL: A suitable tool for HIV-1 and HIV-2 viral load quantification in West Africa

J Virol Methods. 2024 Feb:324:114872. doi: 10.1016/j.jviromet.2023.114872. Epub 2023 Dec 19.

Abstract

Point-of-Care for HIV viral RNA quantification seems to be a complementary strategy to the existing conventional systems. This study evaluated the performance of the m-PIMA™ HIV1/2 Viral Load for the quantification of both HIV-1 and HIV-2 RNA viral load. A total of 555 HIV-1 and 90 HIV-2 samples previously tested by Abbott RealTime HIV-1 (Abbott, Chicago, USA) and Generic HIV-2® Charge virale (Biocentric, France) were tested using the m-PIMA™ HIV1/2 Viral Load at the HIV National Reference lab in Senegal. For HIV-1, Pearson correlation and Bland-Altman plots showed a coefficient r = 0.97 and a bias of -0.11 log10 copies/ml (95% confidence interval [CI]: -0.086 to -0.133 log10 copies/ml) for the m-PIMA™ HIV1/2 Viral Load, respectively. Sensitivity and specificity at 3 log10 copies/ml (threshold of virological failure) were 93.6% (95%[CI]: 91.5% to 95.6%) and 99.1% (95%[CI]: 98.3% to 99.9%), respectively. For HIV-2, a correlation of r = 0.95 was also noted with a bias of - 0.229 log10 copies/ml (95%[CI]: -0.161 to -0.297 log10 copies/ml). Sensitivity and specificity at 3 log10 copies/ml were 97.6% (95%[CI]: 94.3% to 100%) and 93.9% (95%[CI]: 88.9% to 98.8%), respectively. These results confirmed that m-PIMA™ HIV1/2 VL could be a good alternative for HIV-1 and HIV-2 viral load testing in decentralized settings in Senegal.

Keywords: HIV-1; HIV-2; POC; Plasma Viral load; Senegal.

MeSH terms

  • Africa, Western
  • HIV Infections* / diagnosis
  • HIV-1* / genetics
  • HIV-2 / genetics
  • Humans
  • RNA, Viral / genetics
  • Sensitivity and Specificity
  • Viral Load / methods

Substances

  • RNA, Viral