Usefulness and analytical performances of complement multiplex assay for measuring complement biomarkers in plasma

Clin Chim Acta. 2024 Feb 1:554:117750. doi: 10.1016/j.cca.2023.117750. Epub 2024 Jan 2.

Abstract

Introduction: The complement system is involved in numerous diseases, through diverse mechanisms and degree of activation. With the emergence of complement targeting therapeutic, simple and accessible tools to evaluate the extent of complement activation are strongly needed.

Methods: We evaluated two multiplex panels, measuring complement activation fragments (C4a, C3a, C5a, Bb, Ba, sC5b9) and intact components or regulators (C1q, C2, C3, C4, C5, FD, FP, FH, FI). The specificity of each measurement was assessed by using complement proteins depleted sera and plasma collected from patients with complement deficiencies. Normal values distribution was estimated using 124 plasma samples from healthy donors and complement activation profile was assessed in plasma collected from 31 patients with various complement-mediated disorders.

Results: We observed good inter-assay variation. All tested protein deficiencies were accurately detected. We established assay-specific reference values for each analyte. Except for C3, C4 and C4a, the majority of the measurements were in good agreement with references methods or published data.

Conclusion: Our study substantiates the utility of the Complement Multiplex assay as a tool for measuring complement activation and deficiencies. Quantifying complement cleavage fragments in patients exhibiting classical or alternative pathway activation allowed evaluating the activation state of the whole cascade.

Keywords: Biomarkers; Complement; ELISA; Therapeutic.

MeSH terms

  • Biomarkers
  • Complement Activation*
  • Complement System Proteins*
  • Humans
  • Plasma

Substances

  • Complement System Proteins
  • Biomarkers