[Translated article] Experience in major complications with total trapezometacarpal prostheses

Rev Esp Cir Ortop Traumatol. 2024 Jan 15:S1888-4415(24)00034-1. doi: 10.1016/j.recot.2024.01.008. Online ahead of print.
[Article in English, Spanish]

Abstract

Introduction: The treatment of rhizarthrosis using trapeziometacarpal prostheses (TMP) is increasing. Complications may lead to loss of the implant and result in salvage surgery. Our aim was to assess major complications with the use of some TMP models and their rescue.

Material and method: Retrospective study on TMP implanted between 2006 and 2021. Models studied: Arpe®, Elektra®, Ivory®, Maïa®, Isis® and Touch®. Demographic data were assessed, implant placement by radiographic study, technical data, complications, salvage surgeries and final survival.

Results: Review of 224 TMP, 45 Arpe® (95.5% survival, rate follow-up [R] 6-16 years), 5 Elektra® (80% survival, R 13-14), 14 Ivory® (92.8% survival, R 9-11), 7 Maïa® (100% survival, R 8-9), 115 Isis® (99.1% survival, R 1-8), 38 Touch® (100% survival, R 1-4). The medial angle of the dome with the proximal articular surface of the trapezium in the lateral plane, was: Arpe®: 8.85°, Elektra®: not assessable, Ivory®: 6.6°, Maïa®: 14.4°, Isis®: 3.8°, and Touch®: 5.95°. The Isis® was placed 100% with scopic guidance presenting a significantly lower angle respect to the medial angle of the dome with the proximal articular surface of the trapezium. As main complications, we observed 3.5% of dislocations and 4% of mobilisations, with the Elektra® being responsible for 47% of these. Nineteen salvage surgeries were performed, with 3% of the implants being lost.

Conclusions: Dislocation and mobilisation are the most observed complications, the Elektra® responsible for almost half of them. Correct placement and implant design appear to be crucial to avoid them in the short and long term.

Keywords: Complicaciones; Complications; Prótesis trapeciometacarpianas; Rhizarthrosis; Rizartrosis; Trapeziometacarpal prosthesis.