[The quality of the roentgen image--a recommendation for quantifying image quality]

Rontgenblatter. 1987 Jan;40(1):18-22.
[Article in German]

Abstract

A proposal is worked out to express x-ray image quality by means of a meaningful and practically useful numerical system. Attennations have been selected in such a manner that whole numbers of Quality Value figures ranging from -10 to +60 result in meaningful associations with radiological indications. Individual parameters are incorporated into the resulting values. A test body for determining Information Sensitivity figures is described that is practical easy built and inexpensive; its most important feature, however, is that it is reproducible and permits international comparisons. This test body provides a direct linking with the Quality Value figures.

Publication types

  • English Abstract

MeSH terms

  • Humans
  • Image Enhancement / standards*
  • Mathematics
  • Models, Anatomic
  • Technology, Radiologic / standards*