Validation of Fully Automated Robust Multicriterial Treatment Planning for Head and Neck Cancer IMPT

Int J Radiat Oncol Biol Phys. 2024 Jul 1;119(3):968-977. doi: 10.1016/j.ijrobp.2023.12.034. Epub 2024 Jan 27.

Abstract

Purpose: Our purpose was to compare robust intensity modulated proton therapy (IMPT) plans, automatically generated with wish-list-based multicriterial optimization as implemented in Erasmus-iCycle, with manually created robust clinical IMPT plans for patients with head and neck cancer.

Methods and materials: Thirty-three patients with head and neck cancer were retrospectively included. All patients were previously treated with a manually created IMPT plan with 7000 cGy dose prescription to the primary tumor (clinical target volume [CTV]7000) and 5425 cGy dose prescription to the bilateral elective volumes (CTV5425). Plans had a 4-beam field configuration and were generated with scenario-based robust optimization (21 scenarios, 3-mm setup error, and ±3% density uncertainty for the CTVs). Three clinical plans were used to configure the Erasmus-iCycle wish-list for automated generation of robust IMPT plans for the other 30 included patients, in line with clinical planning requirements. Automatically and manually generated IMPT plans were compared for (robust) target coverage, organ-at-risk (OAR) doses, and normal tissue complication probabilities (NTCP). No manual fine-tuning of automatically generated plans was performed.

Results: For all automatically generated plans, voxel-wise minimum D98% values for the CTVs were within clinical constraints and similar to manual plans. All investigated OAR parameters were favorable in the automatically generated plans (all P < .001). Median reductions in mean dose to OARs went up to 667 cGy for the inferior pharyngeal constrictor muscle, and median reductions in D0.03cm3 in serial OARs ranged up to 1795 cGy for the spinal cord surface. The observed lower mean dose in parallel OARs resulted in statistically significant lower NTCP for xerostomia (grade ≥2: 34.4% vs 38.0%; grade ≥3: 9.0% vs 10.2%) and dysphagia (grade ≥2: 11.8% vs 15.0%; grade ≥3: 1.8% vs 2.8%).

Conclusions: Erasmus-iCycle was able to produce IMPT dose distributions fully automatically with similar (robust) target coverage and improved OAR doses and NTCPs compared with clinical manual planning, with negligible hands-on planning workload.

Publication types

  • Validation Study

MeSH terms

  • Automation
  • Head and Neck Neoplasms* / diagnostic imaging
  • Head and Neck Neoplasms* / radiotherapy
  • Humans
  • Male
  • Organs at Risk* / diagnostic imaging
  • Organs at Risk* / radiation effects
  • Proton Therapy* / methods
  • Radiotherapy Dosage*
  • Radiotherapy Planning, Computer-Assisted* / methods
  • Radiotherapy Setup Errors / prevention & control
  • Radiotherapy, Intensity-Modulated* / methods
  • Retrospective Studies