Too young for an acquired cardiomyopathy? Cobalt metallosis as a cardiac amyloidosis mimicker

ESC Heart Fail. 2024 Apr;11(2):1236-1241. doi: 10.1002/ehf2.14695. Epub 2024 Jan 29.

Abstract

Metallosis with subsequent cardiac involvement is a possible long-term complication of hip arthroplasty. We report the case of a young female referred to our centre for the suspicion of cardiac amyloidosis presenting with low electrocardiogram voltage, left ventricular hypertrophy, pericardial effusion, and global and longitudinal systolic impairment with apical sparing pattern. Her medical history was remarkable for arthroplasty in the context of congenital hip dysplasia. Two years prior to presentation, she underwent revision surgery for prosthesis malfunction, and tissue metallosis was initially documented. At the current presentation, cobalt metallosis was confirmed, as the circulating cobalt and chromium levels were severely elevated. The accurate diagnosis prompted the removal of the cobalt source with extensive tissue debridement and the use of chelating agents. Reversal of the cardiac abnormalities occurred as the circulating cobalt levels returned to normal.

Keywords: Cardiomyopathy; Cobalt; Metallosis; Prosthetic device; Reversible.

Publication types

  • Case Reports

MeSH terms

  • Arthroplasty, Replacement, Hip*
  • Cardiomyopathies*
  • Cobalt
  • Female
  • Hip Prosthesis*
  • Humans
  • Prosthesis Failure

Substances

  • Cobalt