Comparative Study of Lesions Obtained through Radiofrequency between the Irrigated Ablation Catheter with a Flexible Tip and the Non-Irrigated Catheter in Ex Vivo Porcine Hearts

Biology (Basel). 2024 Feb 19;13(2):132. doi: 10.3390/biology13020132.

Abstract

Background: At the same conditions of delivered power and contact force, open-irrigated radiofrequency ablation catheters are believed to create deeper lesions, while non-irrigated ones produce shallower lesions. This ex vivo study aims to directly compare the lesion dimensions and characteristics of an irrigated ablation catheter with a flexible tip and a non-irrigated solid-tip catheter.

Methods: Radiofrequency lesions were induced on porcine myocardial slabs using both open-tip irrigated and non-irrigated standard 4 mm catheters at three power settings (20 W, 30 W, and 40 W), maintaining a fixed contact force of 10 gr. A lesion assessment was conducted including the lesion depth, depth at the maximum diameter, and lesion surface diameters, with the subsequent calculation of the lesion volume and area being undertaken.

Results: Irrigated catheters produced lesions with significantly higher superficial widths at all power levels (3.8 vs. 4.4 mm at 20 W; 3.9 mm vs. 4.4 mm at 30 W; 3.8 mm vs. 4.5 mm at 40 W; p = 0.001, p = 0.019, p = 0.003, respectively). Non-irrigated catheters resulted in significantly higher superficial areas at all power levels (23 mm2 vs. 18 mm2 at 20 W; 25 mm2 vs. 19 mm2 at 30 W; 26 mm2 vs. 19 mm2 at 40 W; p = 0.001, p = 0.005, p = 0.001, respectively). Irrigated catheters showed significantly higher values of lesion maximum depth at 40 W (4.6 mm vs. 5.5 mm; p = 0.007), while non-irrigated catheters had a significantly higher calculated volume at 20 W (202 µL vs. 134 µL; p = 0.002).

Conclusions: Radiofrequency ablation using an irrigated catheter with a flexible tip has the potential to generate smaller superficial lesion areas compared with those obtained using a non-irrigated catheter.

Keywords: ablation; arrhythmia; irrigated catheter; lesion characteristics; radiofrequency.

Grants and funding

This research received no external funding.