Control-IQ Technology Use in Individuals With High Insulin Requirements: Results From the Multicenter Higher-IQ Trial

J Diabetes Sci Technol. 2024 Mar 5:19322968241234072. doi: 10.1177/19322968241234072. Online ahead of print.

Abstract

Background: Control-IQ technology version 1.5 allows for a wider range of weight and total daily insulin (TDI) entry, in addition to other changes to enhance performance for users with high basal rates. This study evaluated the safety and performance of the updated Control-IQ system for users with basal rates >3 units/h and high TDI in a multicenter, single arm, prospective study.

Methods: Adults with type 1 diabetes (T1D) using continuous subcutaneous insulin infusion (CSII) and at least one basal rate over 3 units/h (N = 34, mean age = 39.9 years, 41.2% female, diabetes duration = 21.8 years) used the t:slim X2 insulin pump with Control-IQ technology version 1.5 for 13 weeks. Primary outcome was safety events (severe hypoglycemia and diabetic ketoacidosis (DKA)). Central laboratory hemoglobin A1c (HbA1c) was measured at system initiation and 13 weeks. Participants continued using glucagon-like peptide-1 (GLP-1) receptor agonists, sodium-glucose transport protein 2 (SGLT-2) inhibitors, or other medications for glycemic control and/or weight loss if on a stable dose.

Results: All 34 participants completed the study. Fifteen participants used a basal rate >3 units/h for all 24 hours of the day. Nine participants used >300 units TDI on at least one day during the study. There were no severe hypoglycemia or DKA events. Time in range 70-180 mg/dL was 64.8% over the 13 weeks, with 1.0% time <70 mg/dL. Hemoglobin A1c decreased from 7.69% at baseline to 6.87% at 13 weeks (-0.82%, P < .001).

Conclusions: Control-IQ technology version 1.5, with wider range of weight and TDI input and enhancements for users with high insulin requirements, was safe in individuals with T1D in this study.

Keywords: Control-IQ; automated insulin delivery; closed-loop; t:slim X2; total daily insulin.