Assessment of Fitbit Charge 4 for sleep stage and heart rate monitoring against polysomnography and during home monitoring in Huntington's disease

J Clin Sleep Med. 2024 Mar 7. doi: 10.5664/jcsm.11098. Online ahead of print.

Abstract

Study objectives: Wearable devices, monitoring sleep stages and heart rate (HR), bring the potential for longitudinal sleep monitoring in patients with neurodegenerative diseases. Sleep quality reduces with disease progression in Huntington's disease (HD). However, the involuntary movements characteristic of HD may affect the accuracy of wrist-worn devices. This study compares sleep stage and heart rate data from the Fitbit Charge 4 (FB) against polysomnography (PSG) in participants with HD.

Methods: Ten participants with manifest HD wore a FB during overnight hospital-based PSG, and for nine of these participants continued to wear the FB for seven nights at home. Sleep stages (30s epochs) and minute-by-minute HR were extracted and compared against PSG data.

Results: FB-estimated total sleep and wake times, and sleep stage times were in good agreement with PSG, with intra-class correlations 0.79-0.96. However, poor agreement was observed for Wake After Sleep Onset, and the number of awakenings. FB detected wake with 68.6±15.5% sensitivity and 93.7±2.5% specificity, rapid eye movement (REM) sleep with high sensitivity and specificity (78.7±31.9%, 95.6±2.3%), and deep sleep with lower sensitivity but high specificity (56.4±28.8%, 95.0±4.8%). FB HR was strongly correlated with PSG, and the mean absolute error between FB and PSG HR data was 1.16 ± 0.42 bpm. At home, longer sleep and shorter wake times were observed compared to hospital data, while percentage sleep stage times were consistent with hospital data.

Conclusions: Results suggest the potential for long-term monitoring of sleep patterns using wrist-worn wearable devices as part of symptom management in HD.

Keywords: PPG; PSG; home monitoring; neurodegeneration; wearable sensor.