Validation of Dynamic 3D MRI for Urodynamics Assessment Using an Anatomically Realistic In Vitro Model of the Bladder

J Biomech Eng. 2024 Jul 1;146(7):071007. doi: 10.1115/1.4065110.

Abstract

Lowery urinary tract symptoms (LUTS) affect a large majority of the aging population. 3D Dynamic MRI shows promise as a noninvasive diagnostic tool that can assess bladder anatomy and function (urodynamics) while overcoming challenges associated with current urodynamic assessment methods. However, validation of this technique remains an unmet need. In this study, an anatomically realistic, bladder-mimicking in vitro flow model was created and used to systematically benchmark 3D dynamic MRI performance using a highly controllable syringe pump. Time-resolved volumes of the synthetic bladder model were obtained during simulated filling and voiding events and used to calculate volumetric flowrate. During MRI acquisitions, pressure during each event was recorded and used to create PV loops for work assessment. Error between control and MRI-derived volume for voiding and filling events exhibited 3.36% and 4.66% differences, respectively. A slight increase in average error was observed for MRI-derived flowrate when compared to the control flowrate (4.90% and 7.67% for voiding and filling, respectively). Overall, average error in segmented volumes increased with decreasing volume flowrate. Pressure drops were observed during voiding. Pressure increased during filling. Enhanced validation of novel 3D MRI urodynamics is achieved by using high-resolution PIV for visualizing and quantifying velocity inside the bladder model, which is not currently possible with 3D Dynamic MRI.

Keywords: bladder urodynamics; dynamic MRI; lowery urinary tract symptoms; particle image velocimetry.

MeSH terms

  • Magnetic Resonance Imaging
  • Urinary Bladder* / diagnostic imaging
  • Urodynamics*