Pelvic Pseudotumor Associated With a Ceramic Bearing Total Hip

J Am Acad Orthop Surg Glob Res Rev. 2024 Mar 21;8(3):e23.00184. doi: 10.5435/JAAOSGlobal-D-23-00184. eCollection 2024 Mar 1.

Abstract

Pseudotumors have been well documented to occur most frequently in metal-metal bearing total hip arthroplasties and less frequently in metal-polyethylene bearings. There are few cases in the literature of pseudotumors occurring in ceramic-ceramic articulations. We report a case of a large pelvic pseudotumor in a patient with a ceramic-ceramic bearing articulation in a 67-year-old man. In addition to the usual investigations, we did a detailed wear analysis of the ceramic implants and an examination of the soft tissues for particulate debris. The detailed wear analysis did show evidence of stripe wear; however, the volumetric wear was within the expected range. Synchrotron imaging identified strontium and zirconium debris arising from the ceramic surfaces. Although association does not mean causation, no other cause for the large pseudotumor could be identified and presumably represents an idiosyncratic reaction to ceramic debris.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Arthroplasty, Replacement, Hip* / adverse effects
  • Ceramics
  • Hip Prosthesis* / adverse effects
  • Humans
  • Male
  • Polyethylene

Substances

  • Polyethylene