Pseudotumors have been well documented to occur most frequently in metal-metal bearing total hip arthroplasties and less frequently in metal-polyethylene bearings. There are few cases in the literature of pseudotumors occurring in ceramic-ceramic articulations. We report a case of a large pelvic pseudotumor in a patient with a ceramic-ceramic bearing articulation in a 67-year-old man. In addition to the usual investigations, we did a detailed wear analysis of the ceramic implants and an examination of the soft tissues for particulate debris. The detailed wear analysis did show evidence of stripe wear; however, the volumetric wear was within the expected range. Synchrotron imaging identified strontium and zirconium debris arising from the ceramic surfaces. Although association does not mean causation, no other cause for the large pseudotumor could be identified and presumably represents an idiosyncratic reaction to ceramic debris.
Copyright © 2024 The Authors. Published by Wolters Kluwer Health, Inc. on behalf of the American Academy of Orthopaedic Surgeons.