A portable system to measure knee extensor spasticity after spinal cord injury

J Neuroeng Rehabil. 2024 Apr 9;21(1):50. doi: 10.1186/s12984-024-01326-9.

Abstract

Background: The pendulum test is a quantitative method used to assess knee extensor spasticity in humans with spinal cord injury (SCI). Yet, the clinical implementation of this method remains limited. The goal of our study was to develop an objective and portable system to assess knee extensor spasticity during the pendulum test using inertial measurement units (IMU).

Methods: Spasticity was quantified by measuring the first swing angle (FSA) using a 3-dimensional optical tracking system (with external markers over the iliotibial band, lateral knee epicondyle, and lateral malleolus) and two wireless IMUs (positioned over the iliotibial band and mid-part of the lower leg) as well as a clinical exam (Modified Ashworth Scale, MAS).

Results: Measurements were taken on separate days to assess test-retest reliability and device agreement in humans with and without SCI. We found no differences between FSA values obtained with the optical tracking system and the IMU-based system in control subjects and individuals with SCI. FSA values from the IMU-based system showed excellent agreement with the optical tracking system in individuals with SCI (ICC > 0.98) and good agreement in controls (ICC > 0.82), excellent test-retest reliability across days in SCI (ICC = 0.93) and good in controls (ICC = 0.87). Notably, FSA values measured by both systems showed a strong association with MAS scores ( ρ ~ -0.8) being decreased in individuals with SCI with higher MAS scores, reflecting the presence of spasticity.

Conclusions: These findings suggest that our new portable IMU-based system provides a robust and flexible alternative to a camera-based optical tracking system to quantify knee extensor spasticity following SCI.

Keywords: Inertial measurement units; Modified Ashworth Scale; Pendulum test; Reliability; Spasticity; Spinal cord injury.

MeSH terms

  • Humans
  • Knee
  • Lower Extremity*
  • Muscle Spasticity / complications
  • Muscle Spasticity / etiology
  • Reproducibility of Results
  • Spinal Cord Injuries* / complications