Transcatheter Pulmonary Valve Replacement With the Harmony Valve in Patients Who Do Not Meet Recommended Oversizing Criteria on the Screening Perimeter Plot

Circ Cardiovasc Interv. 2024 May;17(5):e013889. doi: 10.1161/CIRCINTERVENTIONS.123.013889. Epub 2024 Apr 12.

Abstract

Background: Anatomic selection for Harmony valve implant is determined with the aid of a screening report and perimeter plot (PP) that depicts the perimeter-derived radius along the right ventricular outflow tract (RVOT) and projects device oversizing. The PP provides an estimation of suitability for implant, but its sensitivity as a screening method is unknown. This study was performed to describe anatomic features and outcomes in patients who underwent Harmony TPV25 implant despite a PP that predicted inadequate oversizing.

Methods: We reviewed RVOT anatomic features and measurements in patients who underwent transcatheter pulmonary valve replacement with the Harmony TPV25 device despite a PP that predicted inadequate oversizing.

Results: This study included 22 patients. There were no unsuccessful implants or adverse valve-related events. Anatomic features varied, but all patients fit into 1 of 3 anatomic types characterized by differences in RVOT dimensions. Type 1 patients (n=9) had a long RVOT with a choke point and a wide main pulmonary artery. Type 2 patients (n=6) had a short RVOT that was pyramidal in shape, with no choke point, and extensive main pulmonary artery lengthening/expansion during systole. Type 3 patients (n=7) had a short, bulbous main pulmonary artery with a choke point and an open pulmonary artery bifurcation.

Conclusions: Transcatheter pulmonary valve replacement with the Harmony valve is feasible in some patients whose PP fit analysis predicts inadequate oversizing. All cases in this series fit into 1 of 3 anatomic patterns, which are not identified in the screening report. Implanters must review cases individually to assess the feasibility of the implant.

Keywords: implant; patient; pulmonary artery; pulmonary valve; tetralogy of Fallot.

MeSH terms

  • Adolescent
  • Adult
  • Cardiac Catheterization* / adverse effects
  • Cardiac Catheterization* / instrumentation
  • Child
  • Clinical Decision-Making
  • Female
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Heart Valve Prosthesis Implantation* / instrumentation
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Patient Selection
  • Predictive Value of Tests
  • Prosthesis Design*
  • Pulmonary Valve Insufficiency / diagnostic imaging
  • Pulmonary Valve Insufficiency / physiopathology
  • Pulmonary Valve Insufficiency / surgery
  • Pulmonary Valve* / diagnostic imaging
  • Pulmonary Valve* / physiopathology
  • Pulmonary Valve* / surgery
  • Recovery of Function
  • Retrospective Studies
  • Treatment Outcome
  • Young Adult