A prospective randomized comparative trial of pediatric C-MAC D-blade video laryngoscope with McCoy direct laryngoscope for intubation in children posted for elective surgical procedures under general anesthesia

Paediatr Anaesth. 2024 Apr 29. doi: 10.1111/pan.14911. Online ahead of print.

Abstract

Background: Pediatric airway management requires careful clinical evaluation and experienced execution due to anatomical, physiological, and developmental considerations. Video laryngoscopy in pediatric airways is a developing area of research, with recent data suggesting that video laryngoscopes are better than standard Macintosh blades. Specifically, there is a paucity of literature on the advantages of the C-MAC D-blade compared to the McCoy direct laryngoscope.

Methods: After Ethics Committee approval, 70 American Society of Anesthesiologists physical status 1 and 2 children aged 4-12 years scheduled for elective surgery under general anesthesia were recruited. Patients were randomly allocated to intubation using a C-MAC video laryngoscope size 2 D-blade (Group 1) and a McCoy laryngoscope size 2 blade (Group 2). The Intubation Difficulty Scale (IDS) for ease of intubation was the primary outcome, while Cormack-Lehane grades, duration of laryngoscopy and intubation, hemodynamic responses, and incidence of any airway complications were secondary outcomes.

Results: Both groups were comparable in terms of patient characteristics. The median (IQR) Intubation Difficulty Scale (IDS) score was better but was statistically nonsignificant with C-MAC (0 [0-0] vs. 0 [0-2], p = .055). The glottic views were superior (CL grade I in 32/35 vs. 23/35, p = .002), and the time to best glottic view (6 s [5-7] vs. 8.0 s [6-10], p = .006) was lesser in the C-MAC D-blade group while the total duration of intubation was comparable (20 s [16-22] vs. 18 s [15-22], p = .374). All the patients could be successfully intubated on the first attempt. None of the patients had any complications.

Conclusion: The C-MAC video laryngoscope size 2 D-blade provided faster and better glottic visualization but similar intubation difficulty compared to McCoy size 2 laryngoscope in children. The shorter time to achieve glottic view demonstrated with the C-MAC failed to translate into a shorter total duration of intubation when compared to the McCoy laryngoscope attributable to a pronounced curvature of the D-blade.

Keywords: child; direct laryngoscopy; hemodynamics; intratracheal; intubation; laryngoscopes; pediatric; video laryngoscope.