Optimum Parameters in Ultrasound Coherent Plane Wave Compounding for High LPWV Estimation: Validation on Phantom and Feasibility in 10 Subjects

J Ultrasound Med. 2024 May 3. doi: 10.1002/jum.16469. Online ahead of print.

Abstract

Objectives: The aim of this study is to determine the optimum and fine values of the number and transmission angles of tilted plane waves for coherent plane-wave compounding (CPWC)-based high local pulse wave velocity (LPWV) estimation.

Methods: A Verasonics system incorporating a linear array probe L14-5/38 with 128 elements and a pulsatile pump, CompuFlow1000, were used to acquire radio frequency data of 3, 5, 7, and 9 tilted plane wave sequences with angle intervals from 0° to 12° with a coarse interval increment step of 1°, and the angle intervals from 0° to 2° with a fine interval increment step of 0.25° from a carotid vessel phantom with the LPWV of 13.42 ± 0.90 m/s. The mean value, standard deviation, and coefficients of variation (CV) of the estimated LPWVs were calculated to quantitatively assess the performance of different configurations for CPWC-based LPWV estimation. Ten healthy human subjects of two age groups were recruited to assess the in vivo feasibility of the optimum parameter values.

Results: The CPWC technique with three plane waves (PRF of 12 kHz corresponding to a frame rate of 4000 Hz) with an interval of 0.75° had LPWVs of 13.52 ± 0.08 m/s with the lowest CV of 1.84% on the phantom, and 5.49 ± 1.46 m/s with the lowest CV of 12.35% on 10 subjects.

Conclusions: The optimum parameters determined in this study show the best repeatability of the LPWV measurements with a vessel phantom and 10 healthy subjects, which support further studies on larger datasets for potential applications.

Keywords: coherent plane wave compounding; local pulse wave velocity; optimum parameter values; transit time method; ultrasound.