Development of a Neuropeptide Y-Sensitive Implantable Microelectrode for Continuous Measurements

ACS Sens. 2024 May 24;9(5):2645-2652. doi: 10.1021/acssensors.4c00449. Epub 2024 May 6.

Abstract

In this work, we present the development of the first implantable aptamer-based platinum microelectrode for continuous measurement of a nonelectroactive molecule, neuropeptide Y (NPY). The aptamer immobilization was performed via conjugation chemistry and characterized using cyclic voltammetry before and after the surface modification. The redox label, methylene blue (MB), was attached at the end of the aptamer sequence and characterized using square wave voltammetry (SWV). NPY standard solutions in a three-electrode cell were used to test three aptamers in steady-state measurement using SWV for optimization. The aptamer with the best performance in the steady-state measurements was chosen, and continuous measurements were performed in a flow cell system using intermittent pulse amperometry. Dynamic measurements were compared against confounding and similar peptides such as pancreatic polypeptide and peptide YY, as well as somatostatin to determine the selectivity in the same modified microelectrode. Our Pt-microelectrode aptamer-based NPY biosensor provides signals 10 times higher for NPY compared to the confounding molecules. This proof-of-concept shows the first potential implantable microelectrode that is selectively sensitive to NPY concentration changes.

Keywords: aptamer biosensor; dynamic measurement; intermittent pulse amperometry; neuropeptide Y; square wave voltammetry.

MeSH terms

  • Aptamers, Nucleotide* / chemistry
  • Biosensing Techniques* / methods
  • Electrochemical Techniques / instrumentation
  • Electrochemical Techniques / methods
  • Microelectrodes*
  • Neuropeptide Y* / analysis
  • Platinum* / chemistry

Substances

  • Neuropeptide Y
  • Platinum
  • Aptamers, Nucleotide