Accuracy of 14 intraocular lens power calculation formulas in extremely long eyes

Graefes Arch Clin Exp Ophthalmol. 2024 May 17. doi: 10.1007/s00417-024-06506-w. Online ahead of print.

Abstract

Purpose: To compare the accuracy of 14 formulas in calculating intraocular lens (IOL) power in extremely long eyes with axial length (AL) over 30.0 mm.

Methods: In this retrospective study, 211 eyes (211 patients) with ALs > 30.0 mm were successfully treated with cataract surgery without complications. Ocular biometric parameters were obtained from IOLMaster 700. Fourteen formulas were evaluated using the optimized A constants: Barrett Universal II (BUII), Kane, Emmetropia Verifying Optical (EVO) 2.0, PEARL-DGS, T2, SRK/T, Holladay 1, Holladay 2, Haigis and Wang-Koch AL adjusted formulas (SRK/Tmodified-W/K, Holladay 1modified-W/K, Holladay 1NP-modified-W/K, Holladay 2modified-W/K, Holladay 2NP-modified-W/K). The mean prediction error (PE) and standard deviation (SD), mean absolute errors (MAE), median absolute errors (MedAE), and the percentage of prediction errors (PEs) within ± 0.25 D, ± 0.50 D, ± 1.00 D were analyzed.

Results: The Kane formula had the smallest MAE (0.43 D) and MedAE (0.34 D). The highest percentage of PE within ± 0.25 D was for EVO 2.0 (37.91%) and the Holladay 1NP-modified-W/K formulas (37.91%). The Kane formula had the highest percentage of PEs in the range of ± 0.50, ± 0.75, ± 1.00, and ± 2.00 D. There was no significant difference in PEs within ± 0.25, ± 0.50 ± 0.75 and ± 1.00 D between BUII, Kane, EVO 2.0 and Wang-Koch AL adjusted formulas (P > .05) by using Cochran's Q test. The Holladay 2modified-W/K formula has the lowest percentage of hyperopic outcomes (29.38%).

Conclusions: The BUII, Kane, EVO 2.0 and Wang-Koch AL adjusted formulas have comparable accuracy for IOL power calculation in eyes with ALs > 30.0 mm.

Keywords: Calculation; Cataract surgery; Extremely long axial length; Formula; Intraocular lens power.