Mid-term performance of His bundle pacing and usefulness of backup leads

Europace. 2024 Jul 2;26(7):euae168. doi: 10.1093/europace/euae168.

Abstract

Ventricular backup leads may be considered in selected patients with His bundle pacing (HBP), but it remains unknown to what extent this is useful. A total of 184 HBP patients were studied. At last follow-up, 147 (79.9%) patients retained His bundle capture at programmed output. His bundle pacing lead revision was performed in 5/36 (13.9%) patients without a backup lead and in 3/148 (2.0%) patients with a backup lead (P = 0.008). One patient without a backup lead had syncope due to atrial oversensing. Thus, implantation of ventricular backup leads may avoid lead revision and adverse events in selected HBP patients.

Keywords: Adverse events; Backup lead; Conduction system pacing; His bundle pacing; Lead revision.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Bundle of His* / physiopathology
  • Cardiac Pacing, Artificial* / methods
  • Electrodes, Implanted
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pacemaker, Artificial*
  • Retrospective Studies
  • Time Factors
  • Treatment Outcome