Quiescent frame, contrast-enhanced coronary magnetic resonance angiography reconstructed using limited number of physiologic frames from 5D free-running acquisitions

Magn Reson Imaging. 2024 Nov:113:110209. doi: 10.1016/j.mri.2024.07.008. Epub 2024 Jul 5.

Abstract

Background: 5D, free-running imaging resolves sets of 3D whole-heart images in both cardiac and respiratory dimensions. In an application such as coronary imaging when a single, static image is of interest, computationally expensive offline iterative reconstruction is still needed to compute the multiple 3D datasets.

Purpose: Evaluate how the number of physiologic bins included in the reconstruction affects the computational cost and resulting image quality of a single, static volume reconstruction.

Study type: Retrospective.

Subjects: 15 pediatric patients following Ferumoxytol infusion (4 mg/kg).

Field strength/sequence: 1.5 T/Ungated 5D free-running GRE sequence.

Assessment: The raw data of each subject were binned and reconstructed into a 5D (x-y-z-cardiac-respiratory) images. 1, 3, 5, 7, and 9 bins adjacent to both sides of the retrospectively determined cardiac resting phase and 1, 3 bins adjacent to the end-expiration phase are used for limited frame reconstructions. The static volume within each limited reconstruction was compared with the corresponding full 5D reconstruction using the structural similarity index measure (SSIM). A non-linear regression model was used to fit SSIM with the percentage of data used compared to full reconstruction (% data). A linear regression model was used to fit computation time with % raw data used. Coronary artery sharpness is measured on each limited reconstructed images to determine the minimal number of cardiac and respiratory bins needed to preserve image quality.

Statistical tests: The coefficient of determination (R2) is computed for each regression model.

Results: The % of data used in the reconstruction was linearly related to the computational time (R2 = 0.99). The SSIM of the static image from the limited reconstructions is non-linearly related with the % of data used (R2 = 0.80). Over the 15 patients, the model showed SSIM of 0.9 with 18% of data, and SSIM of 0.96 with 30% of data. The coronary artery sharpness of images reconstructed using no less than 5 cardiac and all respiratory phases is not significantly different from the full reconstructed images using all cardiac and respiratory bins.

Data conclusion: Reconstruction using only a limited number of acquired physiological states can linearly reduce the computational cost while preserving similarity to the full reconstruction image. It is suggested to use no less than 5 cardiac and all respiratory phases in the limited reconstruction to best preserve the original quality seen on the full reconstructed images.

Keywords: 5D whole-heart MRI; Coronary MRA; Free-running.

MeSH terms

  • Adolescent
  • Algorithms
  • Child
  • Child, Preschool
  • Contrast Media*
  • Coronary Vessels / diagnostic imaging
  • Female
  • Ferrosoferric Oxide
  • Heart / diagnostic imaging
  • Humans
  • Image Interpretation, Computer-Assisted / methods
  • Image Processing, Computer-Assisted / methods
  • Imaging, Three-Dimensional* / methods
  • Magnetic Resonance Angiography* / methods
  • Male
  • Retrospective Studies

Substances

  • Contrast Media
  • Ferrosoferric Oxide