Evaluation of the DxU 850m Iris automated urine microscopy analyzer for identifying culture-negative urine samples: From a perspective of reducing urine culture number

Diagn Microbiol Infect Dis. 2024 Sep;110(1):116439. doi: 10.1016/j.diagmicrobio.2024.116439. Epub 2024 Jul 11.

Abstract

We evaluated the DxU 850m Iris Urine Microscopy analyzer as a screening tool for excluding negative urine samples (n = 1337). At a cutoff of 103 colony counts·mL-1, sensitivity was 55.1 %, specificity 68.6 %. The DxU 850m Iris does not offer acceptable prediction of culture-negative urine samples at the tested cutoff.

Keywords: Bacteriuria screening; DxU 850 Iris; Urinalysis; Urinary tract infection; Urine culture.