Cardiopulmonary bypass in a pediatric patient with factor XII deficiency

J Extra Corpor Technol. 2024 Sep;56(3):125-127. doi: 10.1051/ject/2024021. Epub 2024 Sep 20.

Abstract

The safe use of cardiopulmonary bypass (CPB) relies upon the ability to administer, monitor, and reverse anticoagulation. Although rare, the factor XII deficient patient creates a challenge for the perfusionist due to resultant complications in monitoring anticoagulation. There have been proposed strategies to aid in monitoring anticoagulation in factor XII deficient patients, however, documentation of successful monitoring during CPB is infrequent. With the use of the Hemochron Signature Elite and ACT + cartridges, CPB in a factor XII deficient 8-month-old was completed with predictable and reliable anticoagulation monitoring. This case report explores the current suggestions for factor XII deficiency management with CPB.

Keywords: Anticoagulation; Cardiopulmonary bypass; Coagulopathy; Deficiency; Factor XII; Pediatrics; Perfusion.

Publication types

  • Case Reports

MeSH terms

  • Cardiopulmonary Bypass* / methods
  • Factor XII Deficiency* / blood
  • Female
  • Humans
  • Infant
  • Male