Efficiency of manual resuscitators at birth

Arch Dis Child. 1986 Mar;61(3):300-2. doi: 10.1136/adc.61.3.300.

Abstract

The effectiveness of five neonatal/paediatric manual resuscitators was assessed in a group of babies born by caesarean section. Results showed that devices incorporating a large volume reservoir produced the greatest tidal volume, while those with smaller volume reservoirs could not be considered satisfactory for routine use during neonatal resuscitation.

MeSH terms

  • Apnea / therapy*
  • Humans
  • Infant, Newborn
  • Respiration, Artificial / instrumentation*
  • Tidal Volume