An inherent strength of hydrogen/deuterium exchange coupled to mass spectrometry (HDX-MS) is its ability to detect the presence of multiple conformational states of a protein, which often manifest as multimodal isotopic envelopes. However, the statistical considerations for accurate analysis of multimodal spectra have yet to be established. Here we outline an unrestrained binomial distribution fitting approach with the corresponding statistical tests to accurately detect and, when possible, deconvolute isotopic distributions that contain multiple subpopulations. The algorithms have been incorporated into an updated version of the freely available software, HX-Express, and validated using known mixtures of peptides deuterated to varying degrees. This approach presents a readily accessible tool to fit and interpret bimodal and trimodal behavior in HDX-MS data for mixed populations, EX1 kinetics, and pulse labeling data.
Keywords: EX1; H/D exchange; HDX-MS; bimodal; deconvolution; deuterium; multimodal.